Wie lagere und verwende ich pH-Pufferlösungen richtig?
After preparing the buffer solution, it should be placed in a glass bottle or a polyethylene bottle (alkaline pH buffer solutions such as pH 9.18, pH 10.01, pH 12.46, etc. should be placed in polyethylene bottles). The bottle cap should be tightly closed and stored at low temperature (5-10 ℃) in the refrigerator. It can generally be used for about six months. If turbidity, mold or precipitation are found, it cannot be continued to be used. When using, prepare several 50mL polyethylene small bottles, pour the buffer solution from the large bottles into the small bottles, and let them stand at ambient temperature for 1-2 hours until the temperature is balanced before use. After use, it should not be poured into the large bottle to avoid contamination. The buffer solution in the bottle can be used for 2-3 days under environmental conditions of>10 Grad. Im Allgemeinen können die Lösungen mit pH 7,00, pH 6,86 und pH 14,00 über längere Zeiträume verwendet werden. Lösungen mit einem pH-Wert von 9,18 und einem pH-Wert von 10,01 neigen aufgrund der Absorption von Kohlendioxid aus der Luft zu pH-Änderungen.
Wozu dient eine pH-Pufferlösung?
Kalibrieren Sie das pH-Meter, bevor Sie den pH-Wert messen.
Um die Genauigkeit eines pH-Meters zu überprüfen, führen Sie beispielsweise nach der Kalibrierung des pH-Meters mit pH 6,86 und pH 14,00 die pH-Elektrode in die pH-9,18-Lösung ein und prüfen Sie, ob der Anzeigewert des Instruments mit dem pH-Wert der Standardlösung übereinstimmt.
Ist eine Neukalibrierung des pH-Meters bei allgemeinen Präzisionsmessungen erforderlich? Nach der Kalibrierung und Verwendung des pH-Meters kann es zu Abweichungen oder Veränderungen kommen. Tauchen Sie die Elektrode daher vor dem Test in eine Standardpufferlösung ein, die der gemessenen Lösung nahe kommt, und bestimmen Sie anhand der Größe des Fehlers, ob eine Neukalibrierung erforderlich ist.